SE-6000 SE-6011 Probes (options)

High Voltage Probes (PMK) (made to order)

Model

Bandwidth (MHz)

Attenuation ratio
1:x

Withstand voltage
(duty 10%)

Withstand voltage
(duty 50%)

DC

Length (m)

Remark

PHV1000-RO

400

100

6 kV (500 ms cycle)

3.5 kV (100 ms cycle)

1 kV

2

CAT I

PHVS1000-RO

400

1000

6 kV (duty 10% 500 ms cycle)

3.5 kV (100 ms cycle)

1 kV

2

CAT I

PHV641-LRO

380

100

4 kV (200 ms cycle)

3.5 kV (60 ms cycle)

3 kV (+AC peak)

1.2

AC rms 2 kV

PHV661-LRO

380

100

6 kV (200 ms cycle)

5 kV (60 ms cycle)

4 kV (+AC peak)

1.2

AC rms 2.8 kV

PHV4002-3-RO

100

1000

40 kV (100 ms cycle)

30 kV (30 ms cycle)

20 kV (+AC peak)

3

AC rms 14 kV

PHV1000-RO
PHV1000-RO

PHV6xx-LRO
PHV6xx-LRO

High Voltage Probes

HV-P60A     DC 50 MHz, DC 60 kV, pulse 80 kV

HV-P60A
*The probe receiver (SK-301) is optional.

Specifications and performance

Input/output ratio.....

2000:1

Input RC.....

1000 MW, 5 pF ± 2 pF DC 50 MHz

HV-P30A     DC to 50 MHz, DC to 30 kV, pulse 40 kV

HV-P30A
*The probe receiver (SK-301) is optional.

Specifications and performance

Input/output ratio.....

1000:1

Input RC.....

100 MW, 5 pF ± 2 pF DC 50 MHz

Current Probes

SS-250     DC 100 MHz, MAX 30 A rms

SS-250, PS-26

Specifications and performance

Frequency bandwidth.....

DC 100 MHz (-3 dB)

Maximum permissible input current

   Maximum continuous input range: 30 A rms
   During discontinuous operation: 50 A peak

Output sensitivity.....

0.1 V/A

   Sensitivity accuracy:

±1.0% of reading ±10 mA in the following range: 0 to 30 A rms, DC, and AC 45 to 66 Hz (probe alone)

±2.0% of reading in the following range: more than 30 A rms, 50 A peak or less, DC, and AC 45 to 66 Hz (probe alone)

Noise.....

2.5 mA rms or less (oscilloscope that has a bandwidth of 20 MHz)

Measurable conductor diameter.....

 

f 5 mm

SS-240A     DC 50 MHz, MAX 30 A rms

SS-240A, PS-26

Specifications and performance

Frequency bandwidth.....

DC 50 MHz (-3 dB)

Maximum permissible input current

   Maximum continuous input range: 30 A rms
   During discontinuous operation: 50 A peak

Output sensitivity.....

0.1 V/A

   Sensitivity accuracy:

±1.0% of reading ±10 mA in the following range: 0 to 30 A rms, DC, and AC 45 to 66 Hz (probe alone)

±2.0% of reading in the following range: more than 30 A rms, 50 A peak or less, DC, and AC 45 to 66 Hz (probe alone)

Noise.....

2.5 mA rms or less (oscilloscope that has a bandwidth of 20 MHz)

Measurable conductor diameter.....

 

f 5 mm

Differential probe (made to order)

IE-1165     DC 200 MHz

IE-1165
*Used in combination with an optional high voltage probe

Specifications and performance

Input section

  Frequency bandwidth.....

DC 200 MHz (-3 dB)

  Maximum permissible input voltage... ±250 Vmax

  Input RC.....

1 MW ±1.5%16 pF ±2 pF

Output section

  Output sensitivity.....

10:1 compared to input (with a 50 W load termination)

  Output potential.....

±2 Vmax (with a 50 W load termination)

  Power supply.....

AC adapter or battery box (AA battery x 4)

 

 

Products list | brand | Model No. | Home